Type 1 water distribution system
System achieves resistivities up to 18.2 MΩ-cm and TOC levels below 10ppb.
Virgin polypropylene water pathways prevent recontamination.
Recirculation pump continuously recirculates water throughout the entire system to maintain water purity and reduce the quantity of wastewater required for rinse-up.
Quarter-turn quick release canisters allow for easy cartridge replacement.
0.2μm absolute final filter removes bacteria and particulates that could alter your results.
Digital resistivity meter is automatically temperature-compensated to 25°C, preventing fluctuating readings due to temperature changes.
Product Water
Water Quality (1) . . . . Type 1 Reagent Grade Water (RGW) per
ASTM D1193, NCCLS-ASC-3, and CAP.
Flow rate (Maximum) (2) . .Type 1 RGW-Filtered
Pressure feed (40 psig inlet min.) . . . . . . . . .60 Hz 2.5 lpm
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 Hz 2.0 lpm
Overall Installed Dimensions
3-module
Width . . . . . . . . . . . . . . . . . . . . . . . .30” (762 mm)
Depth . . . . . . . . . . . . . . . . . . . . . .7-1/2” (191 mm)
Height . . . . . . . . . . . . . . . . . . . . . . . .26” (660 mm)
Operating Weight
3-module . . . . . . . . . . . . . . . . . . . .27.2 kg (54 lbs.)
Electrical Requirements
Voltage and Frequency (Nominal)
115 VAC, 50/60 Hz . . . . .98-127 VAC, 47-63 Hz, 1 phase
Resistivity Measurement
Range . . . . . . . . . . . . . . . . . . . . .0.01-18.3 megohm-cm [temperature
compensated to 25°C (77°F.)]
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . .± 3% FS
Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1 constant